Skin / Manual
- 1” bright skin temp. display
- Set temp. display
- % of heater output display on Bar Graph
- Selection of Skin & Manual modes
- Temperature display in °C or °F
- APGAR Timer
- Feed timer
- Audio Alarm mute facility
- Mute time programmable
Skin / Air / Manual
- 1” bright skin temp. display
- 1” bright air temp. display
- % of heater output display on Bar Graph
- Separate Set temp. display
- Temperature display in °C or °F
- APGAR Timer
- Feed Timer
- Baby attending alarm
- Mute time programmable
- Audio Alarm mute facility
Manual Mode
- Time programmable
- Heater output adjustable
- Over temp. cutoff.
Manual Mode
- Time programmable
- Heater output adjustable
- Over temp. cutoff.
- Safety alarms (with status LED) (Actual temp. starts blinking)
Skin Mode Safety alarms (with status LED)
- Safety alarms
- (with status LED)
- Temp. High (>1°C of set temp.)
- Temp. Low (<1°C of set temp.) Temp. Over Range (>38°C)
- Temp. Under Range (<34°C)
- (Actual temp. starts blinking)
- Probe Fail
- Power failure
- Heater Failure
- System Failure
- Safety high temp. cutoff at 38°C
Skin Mode
- Skin Temp. High (<1.5°C of set temp.)
- Skin Temp. Low (<1.5°C of set temp.)
- Skin Temp. Over Range (<38°C)
- Skin Temp. Under Range (<34°C)
- Skin Probe Fail
Air Mode
- Air Temp. High (<1.5°C of set temp.)
- Air Temp. Low (<3.0°C of set temp.)
- Air Temp. Over Range (<39°C)
- Air Probe Fail
Safety high temp. cutoff at
- Skin Temp.< 38°C
- Air Temp.< 39°C
- Power Fail
- Heater Fail
- System Fail
UNDER SURFACE LED PHOTOTHERAPY :
Detachable unit can be used along with Infant Care Trolley, Radiant Warmer.
- 17 High power LEDs emitting blue radiation with a peak at 450-470nm specially used for hyperbilirubinaemia treatment and 4 White LEDs for Observation
- Irradiance more than 25 μwatts/cm2/nm
- Time Totaliser (Dose timing, Dose received, Usage, etc), Cooling fan